Home About us Editorial board Search Ahead of print Current issue Archives Submit article Instructions Subscribe Contacts 218
ORIGINAL ARTICLE
Year : 2017  |  Volume : 8  |  Issue : 1  |  Page : 42-46

Establishing the reference change values (RCVs) and validating the delta check auto-verification in a clinical biochemistry laboratory


1 Department of Biochemistry, Father Muller Medical College, Mangaluru, Karnataka, India
2 St. Aloysius College Mangalore, Mangaluru, Karnataka, India

Correspondence Address:
Denver Clive Fernandez
Department of Biochemistry, Father Muller Medical College, Mangaluru, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0975-9727.199363

Rights and Permissions

Aims: Establishing the reference change values (RCVs) and validating the delta check auto-verification in the hospital information system (HIS). Materials and Methods: This study was conducted in the Hospital Laboratory-Biochemistry. Fifty-one parameters were analyzed in three phases. Phase I: Delta check reference change values were established. Phase II: Delta check auto-verification was validated in the hospital information system. Phase III: Calculation of test requiring manual verification, true and false positive rates. Results: Out of all the test results, 1.35% failed the RCV-delta check thus requiring manual verification, and the remaining 98.65% were auto-verified. Only 0.12% test results were true positives indicating laboratory error, and 1.23% were false positives and were correlated clinically. Ten percent simulated data results and 0.37% actual patient results were not identified by the newly introduced HIS. Conclusions: RCV-delta check is a refined form of the delta checks used to analyze acceptable analytical and biological variation in laboratories. Majority of tests passed the RCV-delta check auto-verification, implying that very few test reports require manual verification. True positives can be detected in the laboratory. All HISs should be validated before implementing complete auto-verification.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed5962    
    Printed248    
    Emailed0    
    PDF Downloaded839    
    Comments [Add]    
    Cited by others 3    

Recommend this journal